BPHASIC DEFIBRILATOR
MEASURING PARAMETERS
CHANNEL 2 CHANNEL
ELECTRODE 3/5 LEADS
ECG Acquistion patient cable and paddles
Display 2 waveforms with HR
and joules selected
Lead lead I, II, III, aVR , aVL ,aVF, v
Gain Selection 5/10/15/20/mm/sec
Sweep Speed 12.5,25 or 50 mm/sec
Heart rate range 30-240 bpm
Heart rate Accuracy +2 bpm
Frequency Response 0.05to 40 Hz
CMRR > 100 dB
Input Impedance >100 M OHMS
Recorder
TYPE High Resolution thermal array recorder
Printing speed 25 mm/sec
printer paper size 57 mm x20 Mtr
details on print HR,ECG w/f along with energy delivery,
Date, Time, Patient Details
(ID ,name, age, Sex),joules selected,
sweep speed and gain
Display
Type Colour TFT display
Size 7 ” Diagonally
Resolution 80 x 480 pixels
Electrical
POWER INPUT AC 230 V,50 HZ
BATTERY RATING 14.8 V-440 MAH,LI-POLYMER
CHARGING TIME 8 HOURS
OPERATING TIME 40-50 Shocks (270 joules) or
more than 3 Hours monitoring
without printing
General
weight 5.4 kg
Dimensions 403(L) x 152 (H) x 324 (w) mm
Working temperature 0-40 deg.cel. & up to
and Humidity 90 % non condensing