HeartStart OnSite Defibrillator AED

Product Code: AVI-1537 Category:

Defibrillator family                                             HS1. Order M5066A

Standard configuration                                      Defibrillator, battery, adult SMART Pads cartridge (1 set), Setup and Maintenance Guides, Owner’s Manual, Quick Reference Guide, date sticker

HeartStart OnSite Ready-Pack configuration   Order option R01. Defibrillator, battery, carry case,                                             adult SMART Pads (1 pre-installed set, 1 spare set), Setup and Maintenance Guides, Owner’sManual, Quick Reference Guide, date sticker

Waveform        Truncated Exponential Biphasic; waveform parameters adjusted as a function of each patient’s impedance

Therapy             Adult defibrillation: peak current 32 A 150 J nominal into a 50-ohm load) Pediatric defibrillation with optional Infant/Child SMART Pads cartridge installed: peak current 19 A (50 J nominal into 50-ohm load

Shock-to-shock cycle time      Typically less than 20 seconds between shocks in a series

Voice instructions                     Detailed voice messages guide responder  through use of the defibrillator

Shock-to-shock cycle time     Typically less than 20 seconds between shocks in a series

CPR coaching                                Instructions for adult or infant/child CPR

available at user’s option

Shock delivery                                                  Via adhesive pads placed on patient’s bare skin

as illustrated on pads

Controls                                                            Green SMART Pads cartridge handle, green On/Off

button, blue i-button, orange Shock button

Indicators                                                          Ready light; blue i-button; caution light,

Shock button lights up when shock is advised 

Physical

Size                                                                  7 cm x 19 cm x 21 cm (2.8˝ x 7.4˝ x 8.3˝) D x H x W

Weight                                                              With battery and pads cartridge: 1.5 kg (3.3 lbs.)

Without battery or pads cartridge: 1 kg (2.4 lbs.) 

Environmental/physical requirements

Sealing                                                             Solid objects per EN60529 class IP2X

Drip-proof per EN60529 class IPX1

Temperature                                                      Operating: 0° – 50° C (32° – 122° F)

Standby: 10° – 43° C (50° – 109° F)

Humidity                                                           Operating: 0% to 95% relative, non-condensing

Standby: 0% to 75% relative, non-condensing

Altitude                                                             Operating: 0 to 15,000 feet

Standby: 0 to 8,500 feet > 48 hours and

8,500 to 15,000 feet < 48 hours

Shock/drop abuse                                             Withstands one-meter drop to any edge,

corner or surface

Vibration                                                           Meets EN1789 random and swept sine,

road ambulance specification in operating

and standby states

EMI (radiated/immunity)                                     Meets EN55011 Group 1 Level B Class B

and EN61000-4-3 

Data recording and transmission

Infrared          Wreless transmission of event data to a Smartphone or PC, using the IrDA protocol

Data stored                  First 15 minutes of ECG and the entire incident’s events and analysis decisions

Patient analysis system 

Patient analysis         Evaluates patient ECG to determine if a rhythm is shockable. Rhythms considered shockable are ventricular fibrillation (VF) and certain ventricular tachycardias (VT) associated with lack of circulation. For safety reasons, some VT rhythms associated with circulation will not be interpreted as shockable, and some very low-amplitude or low-frequency rhythms will not be interpreted as shockable VF.

Quick Shock                       Able to deliver a shock after the end of a CPR interval, typically in 8 seconds

Sensitivity/specificity             Meets AAMI DF80 guidelines and AHA recommendations for adult defibrillation

(Circulation 1997;95:1677-1682)

Artifact detection                   The effects of pacemaker artifact and electrical noise are minimized 

Battery (M5070A)

Type                                                                 9 Volt DC, 4.2 Ah, composed of disposable

long-life lithium manganese dioxide primary cells

Capacity                                                           Minimum 200 shocks or 4 hours of operating

time (EN 60601-2-4:2003)

Install-by date                                                   Battery is labeled with an install-by date of

at least 5 years from date of manufacture

Standby life                                                      Four years typical when battery is installed

by the install-by date (will power the AED

in standby state within the specified standby

temperature range, assuming 1 battery

insertion test and no defibrillation uses) 

SMART Pads 

Adult SMART Pads cartridge                             M5071A defibrillation pads for patients 8 years

of age and older or 25 kg (55 lbs.) and over

Infant/Child SMART Pads cartridge                    M5072A defibrillation pads for patients

under 8 years of age or 25 kg (55 lbs.);

by prescription only

Active surface area                                            85 cm2 (13.2˝2) each

Cable length                                                      Adult SMART Pads: 137.1 cm (54˝)

Infant/Child SMART Pads: 101.6 cm (40˝)

Use-by date                                                      Cartridge is labeled with a use-by date of

at least 2 years from date of manufacture 

Training SMART Pads

M5073A                                                            Adult Training SMART Pads cartridge

M5074A                                                            Infant/Child Training SMART Pads cartridge

Function            Training SMART Pads cartridges feature 8 real-world training scripts; used with trainingmat (included) or with adapters on manikins 

Automated and user-activated self-tests

Daily automatic self-tests                    Tests internal circuitry, waveform delivery system, pads cartridge, and battery capacity

Pads integrity test             Specifically tests readiness-for-use of pads(gel moisture)

Battery insertion test                         Upon battery insertion, extensive automatic self-tests and user-interactive test check

device readiness

Status Indicators                                               Blinking green “Ready” light indicates ready

for use; audible “chirp” indicates need for

maintenance