Features |
Specifications |
|
|
Structure |
Square tubular Electrostatically powder coated steel sections. |
Top Tray |
Steel top tray at eye level to keep monitoring equipments etc. (595 mm °° 330 mm). |
Table Top |
Stainless Steel (595 mm °° 330 mm). |
Drawer |
Extra large one storage drawer. |
Cylinder Yokes |
Pin-indexed gas specific dual oxygen & nitrous oxide. |
Regulators |
Nitrolock system is provided which activates only when oxygen flowmeter control valve opens to prevent the risk of administering 100% nitrous oxide. |
Nitro Lock Unit |
Nitro lock system is provided which activates only when Oxygen Flowmeter control valve opens to prevent the risk of administering 100% Nitrous Oxide. |
Oxygen failure warning Device |
Pneumatic device which activates audible alarm when oxygen supply pressure falls 215 kpa. |
Rotameters |
Long (230 mm) rotating bobbin flowmeters accurately calibrated.
Oxygen – 100 ml/min to 8 Litres/min.
Nitrous Oxide – 200 ml/min to 12 Litres/min.
Oxygen flow control is on left side of flowmeter bank as it is preferred Gas. |
Vaporisers |
Two plenum type vaporiser for Ether and Trilene supplied & one Halothane vaporiser as standard. |
Non-return cum pressure
relief valve |
Minimises risk of backflow of gases, blows off when pressure exceeds 200 cm water coloumn. |
Emergency Oxygen |
Guarded flush button out-let at table level. |
Accessories |
One Schrader type self sealing oxygen out-lets (4.22 kg/cm2) for ventilators etc. Extended rear platform for two 10 litre water capacity cylinders. Circle absorber can be securedly mounted on the left side of the main frame. |
Breathing Attachment |
Standard magill’s circuit |