HeartStart OnSite Defibrillator AED
Defibrillator family HS1. Order M5066A
Standard configuration Defibrillator, battery, adult SMART Pads cartridge (1 set), Setup and Maintenance Guides, Owner’s Manual, Quick Reference Guide, date sticker
HeartStart OnSite Ready-Pack configuration Order option R01. Defibrillator, battery, carry case, adult SMART Pads (1 pre-installed set, 1 spare set), Setup and Maintenance Guides, Owner’sManual, Quick Reference Guide, date sticker
Waveform Truncated Exponential Biphasic; waveform parameters adjusted as a function of each patient’s impedance
Therapy Adult defibrillation: peak current 32 A 150 J nominal into a 50-ohm load) Pediatric defibrillation with optional Infant/Child SMART Pads cartridge installed: peak current 19 A (50 J nominal into 50-ohm load
Shock-to-shock cycle time Typically less than 20 seconds between shocks in a series
Voice instructions Detailed voice messages guide responder through use of the defibrillator
Shock-to-shock cycle time Typically less than 20 seconds between shocks in a series
CPR coaching Instructions for adult or infant/child CPR
available at user’s option
Shock delivery Via adhesive pads placed on patient’s bare skin
as illustrated on pads
Controls Green SMART Pads cartridge handle, green On/Off
button, blue i-button, orange Shock button
Indicators Ready light; blue i-button; caution light,
Shock button lights up when shock is advised
Physical
Size 7 cm x 19 cm x 21 cm (2.8˝ x 7.4˝ x 8.3˝) D x H x W
Weight With battery and pads cartridge: 1.5 kg (3.3 lbs.)
Without battery or pads cartridge: 1 kg (2.4 lbs.)
Environmental/physical requirements
Sealing Solid objects per EN60529 class IP2X
Drip-proof per EN60529 class IPX1
Temperature Operating: 0° – 50° C (32° – 122° F)
Standby: 10° – 43° C (50° – 109° F)
Humidity Operating: 0% to 95% relative, non-condensing
Standby: 0% to 75% relative, non-condensing
Altitude Operating: 0 to 15,000 feet
Standby: 0 to 8,500 feet > 48 hours and
8,500 to 15,000 feet < 48 hours
Shock/drop abuse Withstands one-meter drop to any edge,
corner or surface
Vibration Meets EN1789 random and swept sine,
road ambulance specification in operating
and standby states
EMI (radiated/immunity) Meets EN55011 Group 1 Level B Class B
and EN61000-4-3
Data recording and transmission
Infrared Wreless transmission of event data to a Smartphone or PC, using the IrDA protocol
Data stored First 15 minutes of ECG and the entire incident’s events and analysis decisions
Patient analysis system
Patient analysis Evaluates patient ECG to determine if a rhythm is shockable. Rhythms considered shockable are ventricular fibrillation (VF) and certain ventricular tachycardias (VT) associated with lack of circulation. For safety reasons, some VT rhythms associated with circulation will not be interpreted as shockable, and some very low-amplitude or low-frequency rhythms will not be interpreted as shockable VF.
Quick Shock Able to deliver a shock after the end of a CPR interval, typically in 8 seconds
Sensitivity/specificity Meets AAMI DF80 guidelines and AHA recommendations for adult defibrillation
(Circulation 1997;95:1677-1682)
Artifact detection The effects of pacemaker artifact and electrical noise are minimized
Battery (M5070A)
Type 9 Volt DC, 4.2 Ah, composed of disposable
long-life lithium manganese dioxide primary cells
Capacity Minimum 200 shocks or 4 hours of operating
time (EN 60601-2-4:2003)
Install-by date Battery is labeled with an install-by date of
at least 5 years from date of manufacture
Standby life Four years typical when battery is installed
by the install-by date (will power the AED
in standby state within the specified standby
temperature range, assuming 1 battery
insertion test and no defibrillation uses)
SMART Pads
Adult SMART Pads cartridge M5071A defibrillation pads for patients 8 years
of age and older or 25 kg (55 lbs.) and over
Infant/Child SMART Pads cartridge M5072A defibrillation pads for patients
under 8 years of age or 25 kg (55 lbs.);
by prescription only
Active surface area 85 cm2 (13.2˝2) each
Cable length Adult SMART Pads: 137.1 cm (54˝)
Infant/Child SMART Pads: 101.6 cm (40˝)
Use-by date Cartridge is labeled with a use-by date of
at least 2 years from date of manufacture
Training SMART Pads
M5073A Adult Training SMART Pads cartridge
M5074A Infant/Child Training SMART Pads cartridge
Function Training SMART Pads cartridges feature 8 real-world training scripts; used with trainingmat (included) or with adapters on manikins
Automated and user-activated self-tests
Daily automatic self-tests Tests internal circuitry, waveform delivery system, pads cartridge, and battery capacity
Pads integrity test Specifically tests readiness-for-use of pads(gel moisture)
Battery insertion test Upon battery insertion, extensive automatic self-tests and user-interactive test check
device readiness
Status Indicators Blinking green “Ready” light indicates ready
for use; audible “chirp” indicates need for
maintenance