VENTILATOR AERO V9[ADULT PADIATRIC NEONANTAL]
Display
Display size: 15.6 in diagonal
Display type: active matrix color TFT LCD
Display resolution: SVGA resolution, 1366×768
Controls
Fully electronic control
Ventilation type: blower
Ventilation modes: A/C-VCV. A/C-PCV, A/C-PRVC, SIMV-VC, SIMV-PC, SIMV-PRVC, BIPAP, APRV, CPAP, CPAP-VG, NIV, IV., nCPAP, AMV.
Special Functions
Manual breath, O2 therapy, standby, screen-lock, inspiratory hold, expiratory hold, sigh.ATC, PCVR, PPS, Compliance, p0.1, calibrate, nebulizer, O2 suction.
Patient types: Adult/pediatric/neonate
Patient height: adjustable
Monitoring
Real-time waveforms
Paw. Flow, volume, EtCO2, SPO2
Optional : CO2, SPO2 [related parameter]
Loops : P-V, V-flow, flow-P
Trends : 1s, 5s, 1min, 10min, 30min, 60min
Tidal volume : 2 to 2000 mL
PEEP/CPAP : 0 to 50 cmH2O
Oxygen : 21 to 100%
I:E ratio : 1:1 to 4:1
Inspiratory time [TI] : 0.1 to 10s
Flow trigger : OFF, 1 to 10 L/Mint.
Pressure trigger : -1 to -10 cmH2O
Pressure control : 1 to 99 cm H2O, added to PEEP/CPAP
Pressure support : 1 to 80 cmH2O, added to PEEP/CPAP
Pressure ramp : 0 to 2s
Expiratory trigger sensitivity [ETS]
OFF, 5 to 90% of inspiratory peak flow
Peak flow spontaneous: >210 L/min
Electrical and gas supply
Input voltage: 100 to 120 V ~ + 10%, 50 / 60 Hz AC or 24 VDC + 10%
Power consumption: 50 W typical
Protection class: Class I
Backup battery: 6 hours, 2 batteries [5200 mAh each/hot swappable
Oxygen supply: 280 to 600 kPa [41 to 87 psi], 120 L/min
Low pressure: oxygen < 15L/min, max 600 hPa for low pressure
Alarms
Operator-adjustable: low/high minute volume, low/high pressure, low/high tidal
volume, low/high rate, apnea time, low/high oxygen
Optional: low/high CO2, low/high SPO2